Evolución de la radiología dental intraoral tras la instauración de la nueva legislación de control de calidad

dc.contributor
Universidad de Murcia. Departamento de Dermatología, Estomatología, Radiología y Medicina Física
dc.contributor.author
Martínez Beneyto, Yolanda
dc.date.accessioned
2011-04-12T20:14:28Z
dc.date.available
2006-06-28
dc.date.issued
2003-06-13
dc.date.submitted
2006-06-27
dc.identifier.isbn
8468994359
dc.identifier.uri
http://www.tesisenred.net/TDR-0627106-143850
dc.identifier.uri
http://hdl.handle.net/10803/10753
dc.description.abstract
Se estudian 7.176 informes oficiales de control de calidad de instalaciones de radiodiagnóstico dental pertenecientes a clínicas odontológicas que utilizan algún tipo de aparato de radiología intraoral. El estudio evolutivo corresponde a las cinco primeras revisiones de dichas instalaciones como consecuencia de la nueva normativa en la que se establecen las obligaciones que deben cumplir este tipo de instalaciones (1997-2001). <br/> Los resultados exponen un análisis cuantitativo de los parámetros, recogidos en dichos informes, con especial referencia a las características <br/>fIsicas (kVp, mA. filtración, colimador, longitud del disparador), así como las desviaciones detectadas en su funcionamiento. Se determinan las características del proceso de obtención de las imágenes radiológicas en dichas instalaciones (tipo revelado, control de tiempos de revelado, renovación de líquidos, tipo película radiogrifica, almacenamiento y caducidad de la película) y la dosis media de radiación ionizante empleada en las mismas para obtener la imagen radiológica de una misma pieza dental en las condiciones de trabajo habituales en cada una de las salas. Todo ello en su <br/>proceso evolutivo durante los cinco años posteriores a la instauración de la normativa de control de calidad. <br/><br/> En general, los resultados obtenidos muestran una disminución del 18.75 % en las dosis medias administradas durante estos cinco años de <br/>estudio. El 97.98 % de las instalaciones radiológícas dentales estudiadas emplean dosis iguales o inferiores a 7 mGy para la realización de la exploración analizada y que corresponde a la dosis máxima de referencia recomendada en radiología intraoral. Sin embargo, sólo el 72.79% de las instalaciones radiológícas podrían llegar a cumplir todas las recomendaciones de la Unión Europea sobre las características de funcionamiento de estos equipos. <br/>Aunque el grado de cumplimiento de las recomendaciones oficiales sobre protección radiológíca no es satisfactoria. se han observado mejorías significativas en la actuación radiológíca dental por parte del profesional a lo largo de estos cinco años de evolución. Por ello, la instauración de la normativa legal esta produciendo una mejoría lenta pero progresiva de la actuación dental en España.
spa
dc.description.abstract
' <br/>A total of 7.176 official reports on dental surgeries of cIinics in Spain are analysed. Tbe study covered the period 1997 to 2001 as consecuence of the entry into force of a new Royal Decree and dental surgeries have to undergo an annual control of quality. <br/> Tbe results shows a quantitative análisis of the variables studied in the ~ oficial reports, particu1ar1y physical characteristics of the X-ray equipments (kVp, mA, filtration thickness, collimator, lenth of tube), besides anomalies in their correct way of working. Processing of film characteristics (type of processing, control in processing time, renewal of liquids, type of film, storage and film expiration) and mean dose of ionizaing radiation are determined in habitual conditions. This study have been carried out during the 5 years following the introducction of the EU norm concerning radiological quality control .<br/>In general. the mean dose received by patients fell during the 5 years studied by 18.75 %. Meanwhile, the 97.98 % of equipments fulfil1ed the EU recommendation and did not excede 7 mGy on average per shot. However, only the 72.79 % of dental radiological equipments could follow the UE recommendations on a correct way of working. <br/>Although EU recommendations concerning radiological protection in dental clinics are not met in all cases, it have been detected significant improvements in professional behaviour during this five years of study. The establishment of the new legislation has led a gradual improvement in Spanish dental radiology practices.
spa
dc.format.mimetype
application/pdf
dc.language.iso
spa
dc.publisher
Universidad de Murcia
dc.rights.license
ADVERTENCIA. El acceso a los contenidos de esta tesis doctoral y su utilización debe respetar los derechos de la persona autora. Puede ser utilizada para consulta o estudio personal, así como en actividades o materiales de investigación y docencia en los términos establecidos en el art. 32 del Texto Refundido de la Ley de Propiedad Intelectual (RDL 1/1996). Para otros usos se requiere la autorización previa y expresa de la persona autora. En cualquier caso, en la utilización de sus contenidos se deberá indicar de forma clara el nombre y apellidos de la persona autora y el título de la tesis doctoral. No se autoriza su reproducción u otras formas de explotación efectuadas con fines lucrativos ni su comunicación pública desde un sitio ajeno al servicio TDR. Tampoco se autoriza la presentación de su contenido en una ventana o marco ajeno a TDR (framing). Esta reserva de derechos afecta tanto al contenido de la tesis como a sus resúmenes e índices.
dc.source
TDR (Tesis Doctorales en Red)
dc.subject
Quality assurance
dc.subject
Radiological protection
dc.subject
Película radiográfica dental
dc.subject
Radiología dental
dc.subject
Control de calidad
dc.subject
Protección radiológica
dc.subject
Dental Radiology
dc.subject
Dental film.
dc.subject.other
Estomatología
dc.title
Evolución de la radiología dental intraoral tras la instauración de la nueva legislación de control de calidad
dc.type
info:eu-repo/semantics/doctoralThesis
dc.type
info:eu-repo/semantics/publishedVersion
dc.subject.udc
6
spa
dc.subject.udc
61
spa
dc.subject.udc
615
spa
dc.contributor.authoremail
yolandam@um.es
dc.contributor.director
Alcaraz Baños, Miguel
dc.rights.accessLevel
info:eu-repo/semantics/openAccess
cat
dc.identifier.dl
MU-2152-2006


Documents

MartinezBeneyto.pdf

6.845Mb PDF

This item appears in the following Collection(s)